VNA Commands [:SENSe]:CORRection:COLLect Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-121
Calibration Steps and Calibration Types (continued):
Table 3-9. SSLT Calibration Method
Calibration Type SHORT1 SHORT2 LOAD THRU ISOL
RF2P (Full 2 Port – S
11
, S
21
, S
12
, S
22
) 1,2 1,2 1,2 3 3
RFP1 (Full S
11
- Port 1) 1 1 1 — —
RFP2 (Full S
22
- Port 2) 2 2 2 — —
RFBP (Full S
11
& S
22
- Both Ports) 1,2 1,2 1,2 — —
TRFP (Response S
21
- Trans Response Fwd
Path)
— — —11
TRRP (Response S
12
- Trans Response Rev
Path)
— — —22
TRBP (Response S
21
& S
12
- Trans Resp Both
Paths)
— — —33
RRP1 (Response S
11
- Reflection Response
Port 1)
1 —1— —
RRP2 (Response S
22
- Reflection Response
Port 2)
2 —2— —
RRBP (Response S
11
& S
22
- Reflection
Response Both Ports)
1,2 —1,2— —
2PFP (1P2P S
11
, S
21
- 1 Path 2 Port Fwd Path) 1 1 1 1 1
2PRP (1P2P S
22
, S
12
- 1 Path 2 Port Rev Path) 2 2 2 2 2