:CALCulate<Tr>:FILTer[:GATE] Subsystem VNA Commands
3-4 PN: 10580-00306 Rev. E MS20xxC PM
3-3 :CALCulate<Tr>:FILTer[:GATE] Subsystem
This subsystem includes commands that allow you to set up the gate configuration.
Gate Coupling State
:CALCulate:FILTer[:GATE]:COUPled[:STATe]
Description: Sets the gate coupling state. Setting the value to ON or to 1 turns on the
gate coupling, which implies that the gate settings for all traces are
identical. Setting the value to OFF or to 0 turns off the gate coupling,
which implies that each trace can have different gate settings. The
query version of this command returns 1 if gate coupling is currently on,
otherwise returns 0 for off.
Syntax: :CALCulate:FILTer[:GATE]:COUPled[:STATe] ON|OFF|1|0
:CALCulate:FILTer[:GATE]:COUPled[:STATe]?
Cmd Parameter: <boolean> ON|OFF|1|0
Query Response: <bNR1> 1|0
Default Value: 1
Example: To set gate coupling to off:
:CALC:FILT:COUP OFF
:CALCulate:FILTer:GATE:COUPled:STATe 0
Front Panel
Access: Shift-8 (System), Application Options, Time Domain, Gate Coupled