VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-41
Add Upper Limit Point
:CALCulate<Tr>:LIMit:UPPer:POINt:ADD
Description: Adds a new limit point to the upper limit line of the given trace <Tr>.
<Tr> is the trace number in the range 1 to 4. If no trace number is
specified, then the <Tr> parameter defaults to trace number 1.
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt:ADD
Cmd Parameter: NA
Query Response: NA (no query)
Example: To add a point to the upper limit line on trace 2:
:CALC2:LIM:UPP:POIN:ADD
Related Command: :CALCulate<Tr>:LIMit:UPPer:POINt:DELete
Front Panel
Access: Shift 6 (Limit), Limit Edit, Add Point
Delete Upper Limit Point
:CALCulate<Tr>:LIMit:UPPer:POINt:DELete
Description: Deletes the upper limit point of the given trace <Tr>. After deletion,
the point that is immediately to the left of the point that was deleted
becomes the active point. <Tr> is the trace number in the range 1 to 4.
If no trace number is specified, then the <Tr> parameter defaults to
trace number 1. Note that deletion is valid only if 2 or more limit points
are active.
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt:DELete
Cmd Parameter: NA
Query Response: NA (no query)
Example: To delete trace 3 upper limit current active point:
:CALCulate3:LIMit:UPPer:POINt:DELete
Related Command: :CALCulate<Tr>:LIMit:UPPer:POINt:ADD
Front Panel
Access: Shift 6 (Limit), Limit Edit, Delete Point