VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-51
Limit X Value (continued)
Example: To move the active limit to 5000 Hertz (active trace in frequency
domain):
:CALCulate:LIMit:X 5000
OR to 500 MHz:
:CALCulate:LIMit:X 500MHz
To move the active limit to 5 Feet (active trace in distance domain with
current distance unit in feet):
:CALCulate:LIMit:X 5FT
OR to 4 Meter
:CALCulate:LIMit:X 4M
:CALCulate:LIMit:X 4
To move the active limit to 2.5 nanoseconds (active trace in time
domain):
:CALCulate:LIMit:X 2.5
:CALCulate:LIMit:X 2.5ns
To move the active limit point to 25 microseconds (active trace in time
domain):
:CALCulate:LIMit:X 25µs
Related Command: :CALCulate:LIMit:TYPe
:CALCulate:LIMit:Y
[:SENSe]:TRACe<Tr>:DOMain
[:SENSe]:TRACe<Tr>:SELect
Front Panel
Access: Shift 6 (Limit), Limit Edit, Limit X