:MEASure:CHPower:CHPower?
Description: Sets the active measurement to channel power, sets the default
measurement parameters, triggers a new measurement and returns the
channel power result. To measure both channel power and channel
power density, use :MEASure:CHPower? It is a combination of the
commands :CONFigure:CHPower; :READ:CHPower:CHPower? For
a description of the default channel power measurement parameters,
refer to :CONFigure:CHPower. To make a channel power
measurement with settings other than the default values send:
:CONFigure:CHPower
Commands to set desired settings
:READ:CHPower:CHPower?
Syntax: :MEASure:CHPower:CHPower?
Cmd Parameter: NA (query only)
Default Unit: Current amplitude units
Related Command: :MEASure:CHPower?
:MEASure:CHPower:DENSity?
:CONFigure:CHPower