:CALCulate:LIMit Subsystem VNA Commands
3-40 PN: 10580-00306 Rev. E MS20xxC PM
Limit Type
:CALCulate:LIMit:TYPE
Description: Sets the limit line segment type (upper or lower) to be edited. Set the
value to 1 for Lower limit segment and to 0 for Upper limit line
segment. The query version of the command returns a 1 if the lower
limit line is currently active for editing and returns a 0 if the upper
limit line is currently active for editing.
Syntax: :CALCulate:LIMit:TYPE 0|1
:CALCulate:LIMit:TYPE?
Cmd Parameter: <char> 0|1
Query Response: <char> 0|1
Default Value: 0
Example: To set upper limit line active for editing:
:CALCulate:LIMit:TYPE 0
To set lower limit line active for editing:
:CALCulate:LIMit:TYPE 1
Front Panel
Access: Shift 6 (Limit), Limit
Number of Upper Limit Points
:CALCulate<Tr>:LIMit:UPPer:POINt?
Description: Query only. Returns the number of points currently in the upper limit
line of the given trace <Tr>. <Tr> is the trace number in the range
1 to 4. If no trace number is specified, then the <Tr> parameter
defaults to trace number 1.
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt?
Cmd Parameter: NA (query only)
Query Response: <NR1> <integer>
Example: To query for the upper limit line total point on trace #2:
:CALC2:LIM:UPP:POIN?
Front Panel
Access: NA