EasyManua.ls Logo

Anritsu MS2026C - 6-12:Trace Subsystem; Trace Copy; Trace Exchange

Anritsu MS2026C
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectrum Analyzer Commands :TRACe Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-53
6-12 :TRACe Subsystem
This subsystem contains commands related to the transfer of trace data to and from the
instrument.
Trace Copy
:TRACe:COPY
Description: Copies Trace A to either Trace B or Trace C. Copying Trace A to
Trace B is equivalent to pressing the Shift-5 (Trace), Trace B Operations,
AB on the front panel. This stores Trace A into Trace B and turns on
Trace B if it was off. Copying Trace A to Trace C is equivalent to
pressing the Shift-5 (Trace), Trace C Operations, AC on the front
panel. This stores Trace A into Trace C and turns on Trace C if it was
off.
Syntax: :TRACe:COPY TRACE1,TRACE2|TRACE3
Cmd Parameter: <char> TRACE1,TRACE2|TRACE3
Query Response: NA (no query)
Example: To copy Trace A to Trace B:
:TRACe:COPY TRACE1,TRACE2
To copy Trace A to Trace C:
:TRACe:COPY TRACE1,Trace3
Front Panel
Access: NA
Trace Exchange
:TRACe:EXCHange TRACE2,TRACE3
Description: Swaps Trace B and Trace C.
Syntax: :TRACe:EXCHange TRACE2,TRACE3
Cmd Parameter: <char> TRACE2,TRACE3
Query Response: NA (no query)
Front Panel
Access: Shift-5 (Trace), Trace B Operations, BC
Shift-5 (Trace), Trace C Operations, BC

Table of Contents