Spectrum Analyzer Commands [:SENSe] Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-85
Occupied Bandwidth Measurement Method
[:SENSe]:OBWidth:METHod
Description: Sets the method for calculating occupied bandwidth. XDB calculates the
occupied bandwidth based on points that are a specified number of dB
below the carrier. Issue command [:SENSe]:OBWidth:XDB to set the
number of dB to be used. PERCent calculates the occupied bandwidth
based on points a specified percentage of the carrier power below the
carrier. Issue command [:SENSe]:OBWidth:PERCent to set the
percentage to be used.
Syntax: [:SENSe]:OBWidth:METHod XDB|PERCent
[:SENSe]:OBWidth:METHod?
Cmd Parameter: <char> XDB|PERCent
Query Response: <char> XDB|PERCent
Default Value: PERCent
Related Command: [:SENSe]:OBWidth:XDB [:SENSe]:OBWidth:PERCent
Front Panel
Access: Shift-4 (Measure), OCC BW, Method, % Int Pwr|> dBc
Occupied Bandwidth Percent of Power
[:SENSe]:OBWidth:PERCent
Description: This command sets the percentage of carrier power that is used to
measure the occupied bandwidth. This value is used in the
measurement if :SENSe:OBWidth:METHod is set to PERCent.
Syntax: [:SENSe]:OBWidth:PERCent <percentage>
[:SENSe]:OBWidth:PERCent?
Cmd Parameter: <percentage>
Query Response: <percentage>
Range: 0% to 100%
Default Value: 99
Default Unit: %
Related Command: [:SENSe]:OBWidth:METHod
Front Panel
Access: Shift-4 (Measure), OCC BW, %