[:SENSe]:CORRection:CKIT:USER Subsystem VNA Commands
3-114 PN: 10580-00306 Rev. E MS20xxC PM
3-25 [:SENSe]:CORRection:CKIT:USER Subsystem
This subsystem contains commands to configure the user device under test (DUT).
DUT User Inductance Coefficient value
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:L0|1|2|3
Description: Sets the DUT inductance value for the specified Short component of the
user-defined SOLT Cal Kit.
Syntax: [:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
L0|1|2|3 <inductance>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
L0|1|2|3?
Cmd Parameter: <NRf> <inductance> (e-12, e-24, e-34, e-42)
Query Response: <NR3> <inductance> (e-12, e-24, e-34, e-42)
Example: To set the DUT inductance L1 for User 1 with line type coax to 5:
:SENS:CORR:CKIT:USER:COAX:SOLT:L1 5
To query the DUT inductance L1 for User 1:
:SENS:CORR:CKIT:USER:COAX:SOLT:L1?
The query response would be: 5.000
Front Panel
Access: Shift-2 (Calibrate), DUT Port Setup, Setup User-Defined, Short
DUT User Capacitance Coefficient value
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:C0|1|2|3
Description: Sets the DUT capacitance value for the specified user.
Syntax: [:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:C0|1|2|3 <capacitance>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:C0|1|2|3?
Cmd Parameter: <NRf> <capacitance> (e-15, e-27, e-36, e-45)
Query Response: <NR3> <capacitance> (e-15, e-27, e-36, e-45)
Example: To set the DUT capacitance #1 for User 1 with line type coax to 5:
:SENS:CORR:CKIT:USER:COAX:SOLT:C1 5
Front Panel
Access: Shift-2 (Calibrate), DUT Port Setup, Setup User-Defined, Open