VNA Commands [:SENSe]:CORRection:CKIT Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-113
Syntax: [:SENSe]:CORRection:CKIT:INFormation? <connector>,
[connector-name string]
Cmd Parameter: NA (query only)
Query Parameter: <char> <connector>, [connector-name string]
Query Response: <block> (returns comma-delimted ASCII format)
Example: To get information for K-Conn(M) with cal Kit TOSLK50A-20:
:SENS:CORR:CKIT:INF? KMAL, "TOSLK50A”
or
:SENS:CORR:CKIT:INF? KMAL, "TOSLK50A-20"
The response for this query when using SOLT COAX:
#6000164OPEN=5.010 mm,SHORT=5.010 mm,
THRU=16.070 mm,C0=4.500 e-15,C1=395.000 e-27,
C2=-20.000 e-36,C3=0.400 e-45,L0=4.000 e-12,
L1=-650.000 e-24,L3=39.000 e-33,L3=-0.640 e-42
Front Panel
Access: Shift-2 (Calibrate), DUT Port Setup, DUT Port 1/2
USR2 User 1 Cal Kit information for the current
calibration method.
Query Response: <block> (returns
comma-delimted ASCII format)
USR3 User 1 Cal Kit information for the current
calibration method.
Query Response: <block> (returns
comma-delimted ASCII format)
USR4 User 1 Cal Kit information for the current
calibration method.
Query Response: <block> (returns
comma-delimted ASCII format)
WAVEGUIDE
<connector> Description

Table of Contents