Spectrum Analyzer Commands :MMEMory Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-47
Recall Measurement
:MMEMory:LOAD:TRACe
Description: The instrument must be in the mode of the saved trace in order to recall
that trace. Use :INSTrument:SELect or :INSTrument:NSELect to
set the mode.
Recalls a previously stored measurement trace from the current storage
location. The saved measurement trace that is to be loaded is specified
by <file name>. <file name> should be enclosed in either single
quotes (‘ ’) or double quotes (“ ”) and should contain a file extension.
Note that the trace specified by <file name> should be available at
the current mass storage location. Use the command :MMEMory:MSIS
to set the current mass storage location. The <integer> parameter is
not currently used, but it must be sent.
Send a numeral 1.
File name extensions:
“.spa” for SPA
“.ia” for Interference Analysis
“.cs” for Channel Scanner
Syntax: :MMEMory:LOAD:TRACe <integer>,<file name>
Cmd Parameter: <integer>, <file name>
Example: To recall trace with filename “trace”:
:MMEMory:LOAD:TRACe 1,”trace.spa”
Related Command: :MMEMory:STORe:TRACe
:MMEMory:STORe:TRACe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Recall Measurement
Shift-7 (File), Recall, Change Type (select file type from list)