Spectrum Analyzer Commands [:SENSe] Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-75
Resolution Bandwidth to Span Ratio
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio
Description: Sets the ratio of the resolution bandwidth to the span for use when the
resolution-bandwidth-to-span coupling is enabled. Note that the front
panel interface sets the inverse ratio: the span to the resolution
bandwidth.
Syntax: [:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio <number>
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio?
Cmd Parameter: <number>
Query Response: <number>
Range: 0.00001 to 1
Default Value: 0.00333
Related Command: [:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO
Front Panel
Access: BW, Span/RBW (note that this is the inverse ratio)
Channel Power Integration Bandwidth
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration
Description: Sets the integration bandwidth for the channel power measurement.
Integration bandwidth must be less than or equal to the frequency span.
Syntax: [:SENSe]:CHPower:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration?
Cmd Parameter: <freq>
Query Response: <freq>
Range: 10 Hz to 9 GHz for MS2036C
10 Hz to 15 GHz for MS2037C
10 Hz to 20 GHz for MS2038C
Default Value: 10.35 MHz
Default Unit: Hz
Related Command: [:SENSe]:FREQuency:SPAN
Front Panel
Access: Shift-4 (Measure), Channel Power, Ch Pwr Width