[:SENSe] Subsystem Spectrum Analyzer Commands
6-86 PN: 10580-00306 Rev. E MS20xxC PM
Occupied Bandwidth Measurement State
[:SENSe]:OBWidth:STATe
Description: Sets the state of the occupied bandwidth measurement, ON or OFF.
When using :CONFigure:OBWidth, the state is automatically set to
ON.
Syntax: [:SENSe]:OBWidth:STATe OFF|ON|0|1
[:SENSe]:OBWidth:STATe?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Related Command: :CONFigure:OBWidth
Front Panel
Access: Shift-4 (Measure), OCC BW, On/Off
Occupied Bandwidth dB Down
[:SENSe]:OBWidth:XDB
Description: This command sets the number of dB below the carrier that is used to
measure the occupied bandwidth. This value is used in the
measurement if :SENSe:OBWidth:METHod is set to XDB.
Syntax: [:SENSe]:OBWidth:XDB <rel ampl>
[:SENSe]:OBWidth:XDB?
Cmd Parameter: <rel ampl>
Query Response: <rel ampl>
Range: 0 dBc to 100 dBc
Default Value: 3dBc
Default Unit: dBc
Related Command: [:SENSe]:OBWidth:METHod
Front Panel
Access: Shift-4 (Measure), OCC BW, dBc

Table of Contents