:CALCulate Subsystem Spectrum Analyzer Commands
6-26 PN: 10580-00306 Rev. E MS20xxC PM
Marker (Maximum) Peak Search Right
:CALCulate:MARKer{1|2|3|4|5|6}:MAXimum:RIGHt
Description: Puts the specified marker on the next highest peak to the right of the
current peak. The next highest peak must be above the peak threshold.
If no point meets that criterion, the marker is set to the last point on the
trace.
Syntax: :CALCulate:MARKer{1|2|3|4|5|6}:MAXimum:RIGHt
Related Command: :CALCulate:MARKer:PEAK:THReshold
Front Panel
Access: Marker, More Peak Options, Next Peak Right
Marker Noise
:CALCulate:MARKer{1|2|3|4|5|6}:NOISe[:STATe]
Description: Turns the marker noise on or off for the specified reference marker.
Note that if counter marker is set to on when setting marker noise to on,
then counter marker is set to off. This command is not valid in zero
span.
Syntax: :CALCulate:MARKer{1|2|3|4|5|6}:NOISe[:STATe] OFF|ON|0|1
:CALCulate:MARKer{1|2|3|4|5|6}:NOISe[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Example: To set reference marker #3 as noise marker:
:CALCulate:MARKer3:NOISe ON
:CALCulate:MARKer3:NOISe 1
Front Panel
Access: Marker, More, Marker Noise