VNA Commands [:SENSe]:CORRection:COLLect Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-131
Calibration Thru Line Length
[SENSe:]CORRection:COLLect:EDELay:DISTance
Description: Set the Calibration Thru Line Length. <line length> is in distance.
Syntax: [SENSe:]CORRection:COLLect:EDELay:DISTance
<line length>
[:SENSe]:CORRection:COLLect:EDELay:DISTance?
Cmd Parameter: <NRf> <line length>
Query Response: <NR3> <line length> (returned in millimeters)
Default Value: 0
Default Unit: Meter (m) when setting, millimeters (mm) for query.
Example: To set to 10 meter:
SENS:CORR:COLL:EDEL:DIST 10m
or
SENS:CORR:COLL:EDEL:DIST 10
To query for line length:
:SENS:CORR:COLL:EDEL:DIST?
The response would be: 10000.00
Front Panel
Access: Shift-2 (Calibrate), Cal Line Setup, Line Length (Air)