Spectrum Analyzer Commands :CALCulate Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-21
Marker Noise
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:NOISe[:STATe]
Description: Turns the delta marker noise on or off. Note that if counter marker is
set to on when setting marker noise to on, then counter marker is set to
off. This command is not valid in zero span. The query version of this
command returns a 1 if the specified delta marker is noise marker and
returns a 0 if not.
Syntax: :CALCulate:MARKer{1|2|3|4|5|6}:DELTa:NOISe[:STATe]
OFF|ON|0|1
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:NOISe[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Example: To turn on marker noise for delta marker #1:
:CALCulate:MARKer1:DELTa:NOISe ON
:CALCulate:MARKer1:DELTa:NOISe 1
:CALCulate:MARKer:DELTa:NOISe 1
:CALCulate:MARKer:DELTa:NOISe:STATe ON
To turn on marker noise for delta marker #2:
:CALCulate:MARKer2:DELTa:NOISe ON
:CALCulate:MARKer2:DELTa:NOISe 1
:CALCulate:MARKer2:DELTa:NOISe:STATe ON
To turn off marker noise #5:
:CALCulate:MARKer5:DELTa:NOISe OFF
:CALCulate:MARKer5:DELTa:NOISe 0
:CALCulate:MARKer5:DELTa:NOISe:STATe OFF
Front Panel
Access: Marker, More, Marker Noise