[:SENSe]:CORRection:COLLect Subsystem VNA Commands
3-132 PN: 10580-00306 Rev. E MS20xxC PM
Calibration Thru Line Length
[SENSe:]CORRection:COLLect:EDELay:TIME
Description: Sets the Calibration Thru Line Delay in units of time.
Syntax: [SENSe:]CORRection:COLLect:EDELay:TIME <line delay>
[:SENSe]:CORRection:COLLect:EDELay:TIME?
Cmd Parameter: <NRf> <line delay>
Query Response: <NR3> <line delay> (time returned in nanoseconds)
Range: –100 ms to +100 ms
Default Value: 0
Default Unit: Seconds (s) when setting, nanoseconds (ns) for query.
Example: To set the line delay to 12 millisecond:
SENS:CORR:COLL:EDEL:TIME 12ms
To query for line delay:
:SENS:CORR:COLL:EDEL:TIME?
The response would be: 12000000.000
Front Panel
Access: Shift-2 (Calibrate), Cal Line Setup, Line Delay
Calibration Interpolation
[:SENSe]:CORRection:COLLect:INTerpolation[:STATe]
Description: Turns the calibration interpolation ON or OFF. Note that interpolation
automatically turns On after a flex cal.
Syntax: [:SENSe]:CORRection:COLLect:INTerpolation[:STATe]
OFF|ON
[:SENSe]:CORRection:COLLect:INTerpolation[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: 0
Example: To set interpolation to ON:
:SENS:CORR:COLLect:INT:STAT ON
or
:CORR:COLLect:INT 1
Front Panel
Access: Shift-2 (Calibrate), Interpolation