Vector Voltmeter Commands [:SENSe]:VVM Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 4-11
4-7 [:SENSe]:VVM Subsystem
This subsystem contains commands pertaining to the Vector Voltmeter mode.
Return Measurement Format
[:SENSe]:VVM:FORMat DB|VSWR|IMPedance
Description: Sets the VVM Return type measurement format. The query format of
the command returns the VVM Return type measurement format.
Query: [:SENSe]:VVM:FORMat?
Syntax: [:SENSe]:VVM:FORMat DB|VSWR|IMPedance
Cmd Parameter: <char> DB|VSWR|IMPedance
Query Response: <char> DB|VSWR|IMP
Default Value: DB
Example: To set the type to VSWR:
:SENSe:VVM:FORMat VSWR
Front Panel
Access: CW, Return Meas Format
Table 4-5. [:SENSe]:VVM Subsystem Commands
Keyword Parameter Data or Units
[:SENSe]
:VVM
:CABLe Refer to “[:SENSe:]:VVM:CABLe Subsystem” on page 4-14.
:FREQuency Refer to “[:SENSe]:VVM:FREQuency Subsystem” on page 4-15
:REFerence Refer to “[:SENSe:]:VVM:REFerence Subsystem” on page 4-16