:CALCulate:TRANsform Subsystem VNA Commands
3-80 PN: 10580-00306 Rev. E MS20xxC PM
Low Pass Mode Response
:CALCulate<Tr>:TRANsform:TIME:LPASs:STIMulus
Description: Sets the response type to be used in the low pass transformation process
for the given trace. The response type is set to either Impulse or Step.
<Tr> is the trace number in the range 1 to 4. If no trace number is
specified, then the <Tr> parameter defaults to trace number 1. The
query version of this command returns the string STEP for step
response and IMP for impulse response.
Syntax: :CALCulate<Tr>:TRANsform:TIME:LPASs:STIMulus
STEP|IMPulse
:CALCulate<Tr>:TRANsform:TIME:LPASs:STIMulus?
Cmd Parameter: <char> STEP|IMPulse
Query Response: <char> STEP|IMP
Default Unit: IMP
Example: To Set trace 1 to Step:
:CALCulate:TRANsform:TIME:LPASs:STIMulus STEP
OR:
:CALC1:TRAN:TIME:LPAS:STIM STEP
Related Command: :CALCulate:TRANsform:TIME:TYPE:AUTO
:CALCulate:TRANsform:TIME:TYPE?
Front Panel
Access: Shift-4 (Measure), Domain Selection, Low Pass Response
(Note that access via this key sequence requires that the domain
processing be set to “Low Pass”.)