VNA Commands [:SENSe]:RFON[:STATe] Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-141
Get Frequency List
:SENSe<Tr>:FREQuency:DATA?
Description: Producers the frequency list in Hz for the given trace. <Tr> is the trace
number in the range 1 to 8 (1 to 4 for Traces TR1 to TR4 and 5 to 8 for
Memory M1 to M4). If no trace number is specified, then the <Tr>
parameter defaults to trace number 1. The response begins with an
ASCII header that specifies the number of data bytes. It appears in the
format #AX, where A is the number of digits in X, and X is the
number of bytes that follow the header. Each frequency point is in
scientific notation and separated by a comma delimiter.
Syntax: :SENSe<Tr>:FREQuency:DATA?
Cmd Parameter: NA (query only)
Query Response: <char> <freq> (returns block data in hertz)
Default Unit: Hz
Related Command: :CALCulate<Tr>:TRANsform:TIME:DATA?
:CALCulate<Tr>:TRANsform:DISTance:DATA?
Front Panel
Access: NA
3-28 [:SENSe]:RFON[:STATe] Subsystem
This subsystem sets the state of the RF output signal at the VNA ports.
RF Power In Hold State
[:SENSe]:RFON[:STATe]
Description: Sets the state of the RF output signal at the VNA ports to be either ON
or OFF when the sweep is set to Hold. When set to ON, the RF signal
continues to be energized when the sweep is in hold. When set to OFF,
the RF signal is turned off during the hold condition. Note that if the RF
has been turned OFF during hold, then the sweep may require more
time to stabilize when it is set to run.
Syntax: [:SENSe]:RFON[:STATe] ON|OFF|1|0
Cmd Parameter: <boolean> ON|OFF|1|0
Query Response: <bNR1> 1|0
Default Value: ON
Related Command: [:SENSe]:SWEep:TYPE
:INITiate:HOLD
Front Panel
Access: Shift 3 (Sweep), RF Pwr in Hold