[:SENSe]:CORRection:CKIT:USER Subsystem VNA Commands
3-116 PN: 10580-00306 Rev. E MS20xxC PM
DUT User Open Offset
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide:SOLT
:OPEN
Description: Sets the DUT capacitance value for the specified user.
Syntax: [:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:OPEN <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:OPEN?
Cmd Parameter: <NRf> <length> (millimeters)
Query Response: <NR3> <length> (millimeters)
Example: To set the DUT open offset for User 1 with line type coax to 3 mm:
:SENS:CORR:CKIT:USER:COAX:SOLT:OPEN 3
Front Panel
Access: NA
DUT User Short Offset (SSLT)
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSLT:SHORt[1]|2
Description: Sets the DUT Short offset for the specified user.
Syntax: [:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSLT:SHORt[1]|2 <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSLT:SHORt[1]|2?
Cmd Parameter: <NRf> <length> (millimeters)
Query Response: <NR3> <length> (millimeters)
Example: To set the DUT short offset 2 for User 1 with line type coax and
calibration method SSLT to 3 mm:
:SENSe:CORRection:CKIT:USER:COAX:SSLT:SHORt2 3
Front Panel
Access: NA