[:SENSe]:CORRection:COLLect Subsystem VNA Commands
3-128 PN: 10580-00306 Rev. E MS20xxC PM
Syntax: [:SENSe]:CORRection:COLLect:CONNector<port_no>
<connector>, [connector-name string]
[:SENSe]:CORRection:COLLect:CONNector<port_no>?
Cmd Parameter: <char> <connector>, [connector-name string]
Query Response: <char> <connector><(connector-name string)>
(connector is returned in short form only)
Default Value: NMAL (KMAL, if Option 11 is installed)
Example: To set the DUT connector for port 1 to K-Conn(M) Cal Kit: OSLK50.
:SENS:CORR:COLL:CONN1 KMAL, "OSLK50"
or
:SENS:CORR:COLL:CONN1 KMAL
To query for the DUT connector for port 1:
:SENS:CORR:COLL:CONN1?
For the example above the query response is: KMAL
To set the DUT connector for port 1 to K-Conn(M)
Cal Kit: TOSLK50A-20
:SENS:CORR:COLL:CONN1 KMAL, "TOSLK50A"
or
:SENS:CORR:COLL:CONN1 KMAL, "TOSLK50A-20"
The query response would return the following:
KMAL(TOSLK50A-20)
Front Panel
Access: Shift-2 (Calibrate), DUT Port Setup, DUT Port 1/2
WG20 WG20/WR42/R22
Cal Kit: xxUM220 or xxUA42
Query Response string: “WG20”
USR1 Query response string: “USR1”
USR2 Query response string: “USR2”
USR3 Query response string: “USR3”
USR4 Query response string: “USR4”
WAVEGUIDE
<connector> Description