AM/FM/PM Commands [:SENSe] Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 7-21
[:SENSe]:OBWidth:PERCent <percentage>
[:SENSe]:OBWidth:PERCent?
Title: Occupied Bandwidth Percent of Power
Description: This command sets the percentage of carrier power used to measure the
occupied bandwidth. This value is used in the measurement if
:SENSe:OBWidth:METHod is set to PERCent.
Parameter: <percentage>
Default Value: 99
Default Unit: %
Range: 0% to 100%
Related Command: [:SENSe]:OBWidth:METHod
Front Panel
Access: Shift-4 (Measure), RF Spectrum, %
[:SENSe]:OBWidth:XDB <rel ampl>
[:SENSe]:OBWidth:XDB?
Title: Occupied Bandwidth dB Down
Description: This command sets the number of dB below the carrier used to measure
the occupied bandwidth. This value is used in the measurement if
:SENSe:OBWidth:METHod is set to XDB.
Parameter: <rel ampl>
Default Value: 3 dBc
Default Unit: dBc
Range: 0 to 100 dBc
Related Command: [:SENSe]:OBWidth:METHod
Front Panel
Access: Shift-4 (Measure), RF Spectrum, dBc