Optional sensor supply (CAN-x-CI8-x-SUPPLY)
Default ranges: +2.5 V to +24 V
voltage
+2.5 V
+5.0 V
+7.5 V
+10 V
+12 V
+15 V
+24 V
15 V
current
580 mA
580 mA
400 mA
300 mA
250 mA
200 mA
120 mA
190 mA
net power
1.5 W
2.9 W
3.0 W
3.0 W
3.0 W
3.0 W
2.9 W
3.0 W
set globally for all channels of a module
Isolation
standard
optional, upon request
output to case (CHASSIS, PE)
nominal rating: 50 V, test voltage (10 sec):
300 V, not available with option ±15 V
to output voltage reference ground
Accuracy of output voltage
<0.25% (typ.)
<0.5% (max.)
<0.9% (max.)
<1% (max.)
at terminals, no load
25°C; 2.5 V to 24 V
25°C; 2.5 V to 24 V
over entire temperature range
15 V
2,5 V, 10 V, 15 V
12 V, 15 V
24 V