SAW sensor – Parameter
Type 8098 FLOWave L
15.6.4 Deactivating the damping of the liquid temperature values
By default, the liquid temperature values are not damped.
But if the damping of the liquid temperature values is active, do the following to deactivate it:
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Stand.meas.values
→ Temperature
→ Damping The Currentsettings are displayed
→ None The Newsettings are displayed.
→
The damping of the liquid temperature values is inactive.
15.6.5 Activating the monitoring of the liquid temperature
case:
• the display shows "— —".
• the message" Notemperaturesensordetected" is displayed.
Because of a malfunction in the process, the measured liquid temperature value can be too high or too low.
A monitored value can be:
• in the normal operating range.
• in the warning range,
• in the error range.
You can set 4 limit values: 2 error limits and 2 warning limits.
→
of the liquid temperature.
value and if the monitored value increases or decreases.
By default, the monitoring of the liquid temperature and the diagnostics are all disabled.
To activate the monitoring of the liquid temperature, do the following:
→ Go to the CONFIGURATION view.
→ SAWsensor