Maintenance and troubleshooting
Type 8098 FLOWave L
24.9.11 Message "Internal message store overow"
Message
Symboldisplayedinthe
informationbar
Possiblecause The number of messages generated by the device is higher than the
memory capacity.
Whattodo?
→
correct.
24.9.12 Message "No signals from interdigital transducer"
Message
No signals from interdigital transducer
Symboldisplayedinthe
informationbar
-
Possiblecause The sensor is not operating correctly.
Whattodo?
→ Send the complete device back to Bürkert because the sensor must be
replaced.
24.9.13 Message "No temperature sensor detected"
Message
No temperature sensor detected
Symboldisplayedinthe
informationbar
Possiblecause The temperature of the liquid can neither be measured nor compensated.
Whattodo?
→ If the temperature of the liquid must be measured, send the complete
device back to Bürkert because the sensor must be replaced.
24.9.14 Message "Pipe characteristics have changed: check limits
values"
Message
Pipe characteristics have changed: check limits values
Symboldisplayedinthe
informationbar
(symbol changed)
Possiblecause The transmitter has been associated with another sensor.
Whattodo? →
correct.