218
SAW sensor – Parameter
Type 8098 FLOWave L
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Add.meas.values
→ DF
→ Damping The Currentsettings are displayed
→ Low, Medium and High The Newsettings are
displayed.
→
15.12.4 Activating a user-dened damping of the dierentiation factor
values
• on the display,
• on the totalizers,
• on the outputs. The damping set for an analog output comes in addition to the damping of the DF.
By default, the measured DF values are damped with the level None.
→ Low, Medium or High.
→ Or you can set your own damping parameters with the Special damping.
With the Special damping, you can set 2 parameters:
• Responsetime in seconds,
• the Jumpthreshold-
centage, no damping is applied to the second measured value.
To set your own damping parameters of the DF values, do the following:
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Add.meas.values
English