5.8 Device status indicator
The color of the device status indicator gives the following pieces of information:
• Whether device diagnostics are active on not.
• If device diagnostics are active, then the device status indicator shows whether diagnostics events have
been generated or not. If several diagnostics events have been generated, then the device status indicator
shows the diagnostics event with the highest priority. Refer to
Bürkert Communicator software.
Colour
accordingto
NE107
Colourcode
(foraPLC)
Diagnosticsevent
accordingto
NE107
Meaning
Red Failure, error or
fault
Due to a malfunction of the device or its periphery,
the measured values can be incorrect.
Orange 4 Check function
the correct behaviour of the outputs by simulating
measurement values); the output signal is temporarily
invalid (e.g. frozen).
Yellow 3 The ambient conditions or process conditions for the
device are outside the permitted ranges.
Device-internal diagnostics point to problems in the
device or with the process properties.
Blue 2
required
The device continues to measure but a function is
temporarily restricted.
→ Do the required maintenance operation.
Green - Diagnostics are active and no diagnostics event has
been generated.
White 0 - Diagnostics are inactive.