302
Outputs – Maintenance
Type 8098 FLOWave L
20.5 Checking the correct operation of an on/o output or
a threshold output
→
→ Energize the output.
→ Go to the CONFIGURATION view.
→ Outputs
→ Parameter view.
→ MAINTENANCE view.
→ Output2:digital or Output3:digital
→ Test
→ On or O Check if the output is operating correctly.
→
20.6 Checking the correct operation of a frequency output
→
→ Energize the output.
→ Go to the CONFIGURATION view.
→ Outputs
→ Parameter view.
→ MAINTENANCE view.
→ Output2:digital or Output3:digital
→ Test
→ Check if the output is operating correctly.
→
English