SAW sensor – Parameter
Type 8098 FLOWave L
Is the event associated to the device
status Outofspecication?
Yes
The device status indicator is yellow.
The analogue outputs react
depending on the behaviour chosen
for Outofspecication events.

Out
ofspecication
output switches depending on the
event.
Is the event associated to the device
status Maintenancerequired?
Yes
The device status indicator is blue.
The analogue outputs react
depending on the behaviour chosen
for Maintenancerequired events.

Mainte-
nancerequired
output switches depending on the
event.
... continued from the previous page.
No
 
15.14.1 Enabling the diagnostics for special events in the process
By default, all the diagnostics related to the process are disabled.
To enable the diagnostics of a special event related to the process, do the following:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
Diagevents
Process
 
event is a Failure, an Outofspecication, a Maintenancerequired or a Normalstate

The diagnostics on the special event is enabled.
.