212
SAW sensor – Parameter
Type 8098 FLOWave L
15.11.6 Resetting the default values of the error limits, the warning
limits and the hysteresis of each mass totalizer
The default values of the error limits, the warning limits and the hysteresis of the mass totalizers are the
following:





To reset the default values of the error limits, the warning limits and the hysteresis of each mass totalizer, do
the following:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
Stand.meas.values Masstotalizer1 or Masstotalizer2
Limits
Resettodefault

The limit values and the hysteresis value are reset.

15.11.7 Enabling the user to start, stop or reset each mass totalizer
By default, the user is not allowed to start, to stop or to reset a mass totalizer.
To authorize the user to start, to stop or to reset a mass totalizer, do the following:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
Stand.meas.values Masstotalizer1 or Masstotalizer2
Start/Stop/Reset
Enabled

The user is authorized to start, to stop or to reset a mass totalizer.