SAW sensor – Parameter
Type 8098 FLOWave L
15.13.5 Deactivating the damping of the values of the acoustic
transmission
If the damping of the values of the acoustic transmission factor is active, do the following to deactivate it:
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Add.meas.values
→ Acoustictransmissionfactor .
→ Damping The Currentsettings are displayed
→ None The Newsettings are displayed.
→
The damping of the values of the acoustic transmission factor is inactive.
15.13.6 Changing the error limits, the warning limits and the hysteresis
of the acoustic transmission factor
To change the error limits, the warning limits and the hysteresis of the acoustic transmission factor, do the
following:
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Add.meas.values
→ Acoustictransmissionfactor
→ Limits
→ Settings The Currentsettings are displayed
→
→
→
→
→ The Newsettings are displayed.
→
The limit values and the hysteresis value are changed.