Electrical installation
Type 8098 FLOWave L
8.10 Wiring the device through the M20x1.5 cable glands in
stainless steel (device variant with cable glands)
Put only one cable in each cable gland.

.
To open the front of the transmitter, follow the
8.9.
 
screws of the functional earth plate.
2 screws
2. 
3. 
4. 
20 mm
100 mm ±5
8 mm
 
 Do not remove the nut of a stainless steel


 Turn the nut until the stop. If you turn beyond the
stop, the cable gland unscrews from the device
and the device is no longer tight.
8. Remove the blind plug of the cable gland and
store the blind plug in a safe and clean place.
9. Put the cable through the cable gland as shown



Attach each cable to the functional earth plate.
The shield must be in contact with the functional
earth plate.