SAW sensor – Maintenance
Type 8098 FLOWave L


The slope value of the DF is set.
17.17 Setting the oset value of the liquid density
.

Go to the CONFIGURATION view.
SAWsensor
Parameter view.
MAINTENANCE view.
Calibration
Stand.meas.values
Density
Oset
Entervalue



17.18 Setting the slope value of the liquid density
To enter a slope value for the liquid density, do the following:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
MAINTENANCE view.
Calibration
Stand.meas.values
Density