SAW sensor – Maintenance
Type 8098 FLOWave L
17.8.1 Calibrating the K factor by using a teach-in procedure
depending on the volume ow rate

In order that the calibration result is correct, make sure the following conditions are met during the
teach-in procedure:
the liquid temperature is stable,






Go to the CONFIGURATION view.
SAWsensor
Parameter view.
MAINTENANCE view.
Calibration
Stand.meas.values
Flowrate Kfactor
Teach-inbyvolumeow The current K factor is displayed.





Newsettings are displayed.

The new K factor is used.

.
