277
SAW sensor – Maintenance
Type 8098 FLOWave L
Parameter view.
MAINTENANCE view.
Calibration
Add.meas.values
Concentration1 or Concentration2
Oset
Entervalue



17.21 Calibrating the oset value of the liquid concentration

To get a correct calibration result, make sure that the following conditions are met during the teach-in
procedure:
The liquid temperature is stable.

of bubbles.
During the calibration:
The device status indicator is orange, if the operating mode of the device status indicator is set

).

.


Charge the pipe. To avoid bubbles and air in the pipe, make sure it is full of liquid.
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
MAINTENANCE view.
English