SAW sensor – Parameter
Type 8098 FLOWave L
→ Limits
→ Active
→ Yes.
→ 
The monitoring of the liquid density is active and the device status will change depending on the limits
that have been set.
→ 
.
→ 
.
→ To enable the monitoring, i.e. to be informed when the value of the liquid density is outside the normal
.
15.9.6 Deactivating the monitoring of the liquid density
By default, the liquid density values are not monitored.
But if the monitoring of the liquid density is active, do the following to deactivate it:
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Stand.meas.values
→ Density Limits
→ Active
→ No.
→ 
The monitoring of the liquid density is inactive.