Maintenance and troubleshooting
Type 8098 FLOWave L
24.12.3 Message "Not totally lled"
Message
Symboldisplayedinthe
informationbar
Depends on the device status the event is associated to.
Possiblecause
The calibration has failed. The calibration has failed because of one of the
possible causes:
• The sensor may be broken (valid for product variants with DN08).
•
the liquid and measurement is not possible.
-
toring special events that occur in the process, on the sensor or on the
electronics.
Whattodo?
→
→ If the message Nosignalsfrominterdigitaltransducer is generated
simutaneously, then send the product back to Bürkert.
→ If the message Nottotallylled remains, then send the product back to
Bürkert.
24.12.4 Message "Liquid out of range"
Message
Symboldisplayedinthe
informationbar
Depends on the device status the event is associated to.
Possiblecause
The speed of sound in the liquid is out of range.
-
toring special events that occur in the process, on the sensor or on the
electronics.
The density in the liquid is out of range.
Whattodo?
→
.
24.12.5 Message "Unstable ow rate"
Message
Symboldisplayedinthe
informationbar
Depends on the device status the event is associated to.
Possiblecause
-
ments is too high.
-
toring special events that occur in the process, on the sensor or on the
electronics.
Whattodo?
→
pumps and process valves, is correct.