Maintenance and troubleshooting
Type 8098 FLOWave L
24.12.9 Message "Sound conductivity out of range"
Message
Sound conductivity out of range
Symboldisplayedinthe
informationbar
Depends on the device status the event is associated to.
Possiblecause
There are gas bubbles or solid particles in the liquid.
-
toring special events that occur in the process, on the sensor or on the
electronics.
Whattodo?
→ Search for malfunctions in the process.
→
24.12.10 Message "AO1 open loop" or "AO3 open loop"
Message
or
AO3 open loop
Symboldisplayedinthe
informationbar
Depends on the device status the event is associated to.
Possiblecause
There is a connection problem on the related output.
The current measured in the current loop is too low compared to the
-
toring special events that occur in the process, on the sensor or on the
electronics.
Whattodo?
→
24.12.11 Message "AO1 Diag error" or "AO3 Diag error"
Message
or
AO3 Diag error
Symboldisplayedinthe
informationbar
Depends on the device status the event is associated to.
Possiblecause
There is a connection problem on the related output or a high resistance is
detected in the loop.
-
toring special events that occur in the process, on the sensor or on the
electronics.
Whattodo?
→
→
Disabling an analogue output.