332
Maintenance and troubleshooting
Type 8098 FLOWave L
24.10.3 Message "Calibration cancelled"
Message
Calibration cancelled
Symboldisplayedinthe
informationbar
Possiblecause
time of 30 seconds has elapsed.
Whattodo?
→ Do a new calibration and observe the given instructions.
24.10.4 Message "Calibration cancelled, the ow rate is less than 5%
of the full scale"
Message
Symboldisplayedinthe
informationbar
Possiblecause The calibration has failed because
scale.
Whattodo?
→
→ Do a new calibration.
24.10.5 Message "Resulting K factor is less than 0.8 or higher than
1.2"
Message
Symboldisplayedinthe
informationbar
Possiblecause When calibrating the K factor by using a teach-in procedure depending
-
Whattodo?
→ Do a new calibration.
→ Enter a correct reference value.
24.10.6 Message "Resulting oset is higher than 10°C, 18°F"
Message
Symboldisplayedinthe
informationbar
Possiblecause
Whattodo?
→ Do a new calibration.
→ Enter a correct reference value.
English