SAW sensor – Maintenance
Type 8098 FLOWave L
17.8.2 Calibrating the K factor by using a teach-in procedure
depending on a known volume

To calibrate the K factor by using a teach-in procedure depending on a known volume, do the following:
Prepare a tank which capacity you know. To make sure to get an accurate K factor, prepare the recom-
mended volume of liquid given in Table 38 or in Table 39.
 
Diameter of the measurement
tube
 Recommended volume, in litres,
to get an accurate K factor
3/8''  
  28
DN8  33
 
Diameter of the measurement
tube
 Recommended volume, in litres,
to get an accurate K factor
3/4''  
  38
  
2''  
  
3''  
  
  
DN40  
  249
  408
DN80  

Go to the CONFIGURATION view.
SAWsensor
Parameter view.
MAINTENANCE view.
Calibration
Stand.meas.values
Flowrate
English