SAW sensor – Maintenance
Type 8098 FLOWave L
17.8.2 Calibrating the K factor by using a teach-in procedure
depending on a known volume
→
To calibrate the K factor by using a teach-in procedure depending on a known volume, do the following:
→ Prepare a tank which capacity you know. To make sure to get an accurate K factor, prepare the recom-
mended volume of liquid given in Table 38 or in Table 39.
Diameter of the measurement
tube
Recommended volume, in litres,
to get an accurate K factor
3/8''
28
DN8 33
Diameter of the measurement
tube
Recommended volume, in litres,
to get an accurate K factor
3/4''
38
2''
3''
DN40
249
408
DN80
→
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ MAINTENANCE view.
→ Calibration
→ Stand.meas.values
→ Flowrate