196
SAW sensor – Parameter
Type 8098 FLOWave L
15.9.3 Activating a user-dened damping of the liquid density values
• on the display,
• on the outputs. The damping of the liquid density comes in addition to the damping set for each analog
).
•
→ Low, Medium or High.
By default, the measured liquid density values are damped with the level Medium.
→ Or you can set your own damping parameters with the Special damping.
With the Special damping, you can set 2 parameters:
• Responsetime in seconds,
• the Jumpthreshold-
centage, no damping is applied to the second measured value.
To set your own damping parameters of the measured liquid density values, do the following:
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Stand.meas.values
→ Density
→ Damping The Currentsettings are displayed
→ Special
→ Responsetime
→ Jumpthreshold is enabled or disabled
→ Jumpthreshold is enabled, set the value. The Newsettings are displayed.
→
The special damping of the liquid density values is active.
English