SAW sensor – Parameter
Type 8098 FLOWave L
15.15.5 Activating the compensation for a liquid with an inverse
quadratic viscosity compensation curve
To activate the compensation for a liquid with an inverse quadratic compensation curve, do the following:
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ Viscositycompensation
→ Settings The current settings are displayed.
→
→ Inversequadratic
→
→ of the quadratic curve, in the displayed units (mm
2
/s), and in the sci-
→
→
notation.
→
→
notation.
→
→
The compensation for a liquid with an inverse quadratic compensation curve is active.
the range covered by the equation, or if wrong constant values have been entered), the compensated volume
Viscositycompensationfailed is displayed. If the message is dis-
played, do the following:
→
→
English