273
SAW sensor – Maintenance
Type 8098 FLOWave L
17.14 Setting the oset value of the dierentiation factor
.
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ MAINTENANCE view.
→ Calibration
→ Add.meas.values
→ DF
→ Oset
→ Entervalue
→
→
17.15 Calibrating the oset value of the dierentiation factor
→
→ To get a correct calibration result, make sure that the following conditions are met during the teach-in
procedure:
• The liquid temperature is stable.
•
of bubbles.
During the calibration:
• The device status indicator is orange, if the operating mode of the device status indicator is set
").
•
.
→
→ Charge the pipe. To avoid bubbles and air in the pipe, make sure it is full of liquid.
English