SAW sensor – Maintenance
Type 8098 FLOWave L
→ Flowrate Kfactor
→ Teach-inbymassowrate The current K factor is displayed.
→
→
→
→ Newsettings are displayed.
→
The new K factor is used.
simulation.
→
17.8.4 Calibrating the K factor by using a teach-in procedure
depending on a known mass
→
To calibrate the K factor by using a teach-in procedure depending on a known mass, do the following:
→ Prepare a tank which capacity you know. To make sure to get an accurate K factor, prepare the recom-
mended mass of liquid given in Table 40 or in .
Diameter of the measurement
tube
Recommended mass, in kg, to get
an accurate K factor
3/8"
DN8
Diameter of the measurement
tube
Recommended mass, in kg, to get
an accurate K factor
3/4"
2"
3"