SAW sensor – Maintenance
Type 8098 FLOWave L
17.11 Calibrating the oset value of the liquid temperature
→
→ Charge the pipe.
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ MAINTENANCE view.
→ Calibration
→ Stand.meas.values
→ Temperature
→ Oset
→
the usual measuring conditions.
→
→ Temper.cal.byref.
→
→ -
erence temperature sensor.
→ Newsettings are displayed.
→
-
lation. The calibration can fail due to the following causes:
• ±
• the integrated temperature sensor is defective.
→