SAW sensor – Maintenance
Type 8098 FLOWave L
17.11 Calibrating the oset value of the liquid temperature





Charge the pipe.
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
MAINTENANCE view.
Calibration
Stand.meas.values
Temperature
Oset

the usual measuring conditions.

Temper.cal.byref. 

-
erence temperature sensor.
Newsettings are displayed.


-
lation. The calibration can fail due to the following causes:
±
the integrated temperature sensor is defective.
