SAW sensor – Parameter
Type 8098 FLOWave L
15.10.6 Resetting the default values of the error limits, the warning
limits and the hysteresis of each volume totalizer
The default values of the error limits, the warning limits and the hysteresis of the volume totalizers are the
following:

3
,

3
,

3
,

3
,

3
.
To reset the default values of the error limits, the warning limits and the hysteresis of each volume totalizer,
do the following:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
Stand.meas.values Totalizer1 or Totalizer2
Limits
Resettodefault

The limit values and the hysteresis value are reset.

15.10.7 Enabling the user to start, stop or reset each volume totalizer
By default, the user is not allowed to start, to stop or to reset a volume totalizer.
To authorize the user to start, to stop or to reset a volume totalizer, do the following:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
Stand.meas.values Totalizer1 or Totalizer2
Start/Stop/Reset
Enabled

The user is authorized to start, to stop or to reset a volume totalizer.