SAW sensor – Maintenance
Type 8098 FLOWave L
17.4 Setting the direction of the ow
rate values are negative.
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ MAINTENANCE view.
→ Flowdirection
→ Standard
Reverse
→
17.5 Calibrating the oset value of the ow zero point
Adjust this parameter:
• before carrying out a teach-in procedure of the K factor.
• after maintenance work.
•
During the calibration:
• The device status indicator is orange, if the operating mode of the device status indicator is set
).
•
.
→ Charge the pipe. To avoid bubbles and air in the pipe, make sure it is full of liquid.
→ Go to the CONFIGURATION view.
→ SAWsensor
→ Parameter view.
→ MAINTENANCE view.
→ Calibration