SAW sensor – Parameter
Type 8098 FLOWave L
15.10.11 Resetting each volume totalizer to a Presetvalue
If the Start/Stop/Reset of a volume totalizer is active, do the following to reset the volume totalizer to the
Presetvalue:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
Stand.meas.values Totalizer1 or Totalizer2
Valuereset

The volume totalizer is reset to the preset value.

volume totalizer reset.
15.10.12 Changing the Presetvalue for a volume totalizer reset
The default value of the Presetvalue is 0.0 l.
If the Start/Stop/Reset of a volume totalizer is active, do the following to change the preset value:
Go to the CONFIGURATION view.
SAWsensor
Parameter view.
Stand.meas.values
Totalizer1 or Totalizer2
Presetvalue


The value is changed.