77
Commissioning
Type 8098 FLOWave L
9.5 First commissioning for detecting bubbles in the pipe
Energise the device.
2. Connect the device to the Bürkert Communicator software. Print a pdf report of all the current settings
Bürkert Communicator software from country.burkert.com.
3. On the display module, do the settings of the Quickstart.
4. If the liquid is not water, then read out the value of the Acoustictransmissionfactor. Refer to
.
•
• liquid might not be measured accurately by
the device.
If the liquid is not water, then read out the value of the DF
•
•
accurately by the device.
Adjust the parameter Damping of the acoustic transmission factor, depending on your application. Refer
or .
Check the correct behaviour of the device by using the menu Simulation.
8. With the Bürkert Communicator software, print a pdf report of the new settings of the device.
9.
Disconnect the Bürkert Communicator software from the device.
English